机构:[1]College of Information Engineering and Automation Kunming University of Science and Technology Kunming, China[2]Centre of PET/CT Kunming General Hospital of PLA Kunming, China[3]Centre of PET/CT First People’s Hospital of Yunnan Province Kunming, China医技片PET/CT云南省第一人民医院[4]Department of medical imaging First People’s Hospital of Yunnan Province Kunming, China医技片磁共振科云南省第一人民医院
Gating technique is a common way to reduce the effect of breathing movement in the imaging of positron emission tomography / computed tomography (PET/CT). Gating threshold value selection will directly affect the quality level of respiratory motion gating. Most methods of gating threshold value selection are implemented by manually determining, which easily leads to an inaccurate gating. Therefore, exploring an effective method of determining the gating threshold value for respiratory motion gating is very important. In this study, a method to automatically determine the threshold value by the slope rate relationship between the adjacent absolute errors of coincidence events was proposed. We use geant4 application tomographic emission (GATE) software package to simulate realistic PET/CT scanner, and build geometry phantom for respiratory motion simulation. Compared with the manually threshold value determination, the proposed method can make the result of gating correction closer to the real situation. The results prove that the method of automatic determination the threshold value can efficiently and automatically determine the threshold value, and overcome the limitation of the manually threshold value determination.
基金:
This study was supported by the National Natural Science Foundation of
China (No.11265007), and Supported by Scientific Research Foundation for
Returned Scholars, Ministry of Education of china (No. 2010-1561).
语种:
外文
WOS:
第一作者:
第一作者机构:[1]College of Information Engineering and Automation Kunming University of Science and Technology Kunming, China
通讯作者:
推荐引用方式(GB/T 7714):
He Jianfeng,Cui Rui,Fu Zeng,et al.Gating Threshold Value Determination for the Motion Gating of Positron Emission Tomography/Computed Tomography[J].2014 7TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2014).2014,189-193.doi:10.1109/BMEI.2014.7002768.
APA:
He, Jianfeng,Cui, Rui,Fu, Zeng,Hong, Yu,Ji, Yunhai...&Wang, Tianchao.(2014).Gating Threshold Value Determination for the Motion Gating of Positron Emission Tomography/Computed Tomography.2014 7TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2014),,
MLA:
He, Jianfeng,et al."Gating Threshold Value Determination for the Motion Gating of Positron Emission Tomography/Computed Tomography".2014 7TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2014) .(2014):189-193